To avoid the cross contamination, the endoscope ought to be disinfected after using. But how to accurately evaluate the disinfection efficacy, it has been a difficult problem for a long time.
TAILIN innovatively developed the endoscope microbial limit test system to solve this problem.
The Endoscope Microbial Test System consists of HTY-WP02 peristaltic pump for sampling,HTY-101 diaphragm pump for suction filtering, and FC501 filtration cup.
Step 1: Rinse the endoscope with sterile water, and transferring the rinsed water in FC501 filtration cup by HTY-WP02 peristaltic pump;
Step 2: Install the FC501 on the HTY-101 diaphragm pump;
Step 3: Start the HTY-101 diaphragm pump, the sampling rinsed water will filtering through the FC501 filtration cup.
Step 4: The microorganism will be hold on the filtration membrane, if there is any.
Step 5: Put the filtration membrane in the culture media for incubation and counting.
The Endoscope Microbial System can fit for most of endoscope brands; it is closed system which can avoid the external contamination when checking the endoscope disinfection/cleaning efficacy.
Excellent peristaltic pump for stable and reliable operation;
Speed regulation is easy;
Display speed and timing parameters with LED digital tube for easy operation;
Built-in lithium battery, with overcharge, overcurrent, overvoltage protection;
Made of polished mirror stainless steel, easy to clean and disinfect;
Compact size, easy to move.
Power supply: DC 12V/ 2A
Power consumption: 20W
Pumping flow: ≥1200mL/min (at max. speed)
Noise: ≤60dB (load condition)
Net Weight:2.0kgs
Dimensions:11cm x 17cm x 12cm
Consumables: FC501 filtration cup
Used for transferring the rinsed water to filtration cup during endoscope microbial test.
Technical specifications are subject to change without notice. Copyright reserved by Tailin.